Skip to main content

COMPARISON OF MICROSCOPIC OBSERVATION TECHNIQUES TO DETERMINE VICKERS INDENTATION DIAGONAL LENGTH

S. Takagi, Y. Tanaka, Y. Seino

Abstract

In order to investigate the best way to determine the diagonal length of a Vickers indentation, bright-field (BF), dark-field (DF), differential interference contrast (DIC) and confocal scanning microscopy is utilized to observe indentations on various samples. Each of BF, DF and DIC methods has advantage and disadvantage and evaluated values should be considered carefully. 3D feature of indentation evaluated with a confocal microscope gives useful information about it. To eliminate the effect of focusing error, focus stacking technique is quite efficient and expected to reduce the uncertainty of measurement of diagonal length. The roughness of sample surface should be taken into account but not only the amplitude parameter, Ra, but also the wavelength parameter, RSm.

Keywords
Vickers hardness; diagonal length measurement; optical microscopy; focus stacking; confocal scanning microscopy
Download
IMEKO-TC5-2022-049.pdf
DOI
10.21014/tc5-2022.049
IMEKO TC
TC5 - Hardness Measurement

Event details

Event
TC5 Conference 2022
Technical Committee
TC5
Email
imeko2022@fsb.hr
Place
Cavtat-Dubrovnik, CROATIA
Time
11 October 2022 - 13 October 2022
Website
https://conferences.imeko.org/event/1/

Back to the proceedings