LOW FREQUENCY PRIMARY VIBRATION CALIBRATION USING A MULTI-COMPONENT SHAKER
Abstract
The multi-component vibration calibration device at PTB consists of a triaxial shaker which is suitable and convenient for calibrating multi-axial seismometers. Three laser vibrometers are furthermore utilised as references. The frequency range was further extended for a European research project. Instead of a closed-loop control system limiting the vibration frequency down to 0.4 Hz, a signal generator is now used to provide the excitation voltage to the amplifier of the shaker in open-loop mode. This excites the shaker yielding vibrations with significantly lower frequencies. The first test measurements with a seismometer were carried out. The results demonstrate that a calibration down to 10 mHz is easily achievable.
Event details
- Event
- TC22 Conference 2022
- Technical Committee
- TC22
- imeko2022@fsb.hr
- Place
- Cavtat-Dubrovnik, CROATIA
- Time
- 11 October 2022 - 13 October 2022
- Website
- https://conferences.imeko.org/event/1/