Skip to main content

Improvement and Analysis of Low Pressure Gas Flow Calibration System - Bell Prover

Chih-Shan Yen, Ying-Chun Lin, Win-Ti Lin, Sheng-Cyuan Fan

Abstract

Bell Prover is the primary standard system for low-pressure gas by National Measurement Laboratory. When Bell Prover conducts calibration, the density of the gas inside the bell jar must be determined by the measurement of the pressure and temperature of the gas inside the bell jar. It is essential to accurately measure the representative temperature of the bell jar interior. After the system is redesigned in temperature control and temperature measurement methods, the half interval of the measurement error due to temperature stratification inside the bell jar is reduced from the original 0.1 ℃ drops to 0.05 ℃, and the relative expanded uncertainty of the entire system drops from 0.11 % to 0.09 %.

Keywords
Bell Prover; gas temperature measurement; measurement uncertainty
Download
IMEKO-TC9-2019-004.pdf
DOI
10.21014/tc9-2022.004
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2022
Technical Committee
TC9
Email
lich@nim.ac.cn
Place
Chongqing, CHINA
Time
1 November 2022 - 4 November 2022
Website
http://flomeko2022.msmk.tech/c/index.html

Back to the proceedings