Study on the Comparison of Three Typical Gas Flow Standards for Calibrating Low Gas Flowmeter

Biyu Zhu, Chao Chen, Li Ma

Abstract

A gas flowmeter for measuring low flow rate has been widely used in the field of medical, health, environmental protection, energy industry, aerospace, etc. To against Covid-2019, the requirement on the low flow rate has been increasing dramatically. At present, the typical standard devices for calibrating low gas flowmeter mainly include standard bell provers of gas flow, standard piston provers of low gas flow and standard laminar of low gas flow. Different measuring principles are adopted among these typical standard devices. To ensure the consistency of these typical standard devices, a comparison test is performed. The standard devices used in the comparison are of the same accuracy grade, with an extended uncertainty of 0.2 %(k = 2). The piston-type gas flow calibrator of grade 1.0 is selected as the transfer standard, and three flow points with high flow rate, medium flow rate and low flow rate are selected for test. The consistency of measurement results is evaluated by normalized deviation En. The comparison results are acceptable which show that three typical standard devices are accurate and reliable.

Keywords
Low gas flow; Standard device; Comparison
Download
IMEKO-TC9-2019-070.pdf
DOI
10.21014/tc9-2022.070
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2022
Technical Committee
TC9
Email
lich@nim.ac.cn
Place
Chongqing, CHINA
Time
1 November 2022 - 4 November 2022
Website
http://flomeko2022.msmk.tech/c/index.html

Back to the proceedings