Skip to main content

Investigation of the pVTt Gas Flow Standard with Active Thermal Compensation

Hu Wenqing, Xu Zhipeng, Zhang Gaoming, Guo Jing

Abstract

pVTt has been employed as the primary gas flow standard in many countries due to its unique advantages like high accuracy, simple structure, low maintenance, etc. However, its calibration efficiency is relatively low because of the long stabilization time of intake and outtake procedures. To reduce the stabilization time, a novel pVTt gas flow standard with active thermal compensation was proposed. Above all, structure andprinciple of the new pVTt standard device was introduced. Subsequently, the numerical simulation was performed to investigate the influence of thermal compensation. Finally, experiments were carried out, and results showed that the stabilization time after intake completion could be reduced from ten minutes to one minute, which means calibration efficiency was improved a great deal.

Keywords
pVTt; Temperature field characteristic; Numerical simulation; Active compensation
Download
IMEKO-TC9-2019-111.pdf
DOI
10.21014/tc9-2022.111
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2022
Technical Committee
TC9
Email
lich@nim.ac.cn
Place
Chongqing, CHINA
Time
1 November 2022 - 4 November 2022
Website
http://flomeko2022.msmk.tech/c/index.html

Back to the proceedings