Skip to main content

A Novel Multi-Parameter Calibrator for Ventilator Tester Based on Reciprocating Plunger

X. Zheng, X. Zhipeng, J. Qing, Z. Gaoming, T. Jianbing

Abstract

Ventilators are widely needed when the COVID-19 is a global outbreak, and they are used to provide mechanical ventilation for patients who are physically unable to breathe, or breathe insufficiently. A ventilator tester is an instrument used to verify and calibrate ventilation parameters of ventilators like gas flow, tidal volume, frequency. As a measuring device, the ventilator tester also needs to be calibrated periodically. And different parameters are usually calibrated with different devices. In order to improve calibration efficiency and accuracy, a novel multi-parameter calibrator for ventilator tester based on reciprocating plunger is proposed in this work. The system composition is introduced and mathematical models are deduced. According to calibrating regulation, different calibrating modes are simulated and realized.

Keywords
Gas flow rate; Dynamic flow rate; Ventilator; Plunger
Download
IMEKO-TC9-2019-142.pdf
DOI
10.21014/tc9-2022.142
IMEKO TC
TC9 - Flow Measurement

Event details

Event
FLOMEKO 2022
Technical Committee
TC9
Email
lich@nim.ac.cn
Place
Chongqing, CHINA
Time
1 November 2022 - 4 November 2022
Website
http://flomeko2022.msmk.tech/c/index.html

Back to the proceedings