Ionic liquid functionalized graphene oxide decorated by gold nanoparticles for electrochemical detection of pirimicarb pesticides
Abstract
Monitoring pesticide contaminants in diverse environmental systems has become a significant concern. A non-enzymatic electrochemical sensor for pirimicarb pesticide detection has been produced by functionalization of the screen-printed carbon electrodes (SPCE) with graphene oxide, ionic liquid, and gold nanoparticles (GO/IL/AuNps/SPCE). Cyclic voltammetry and electrochemical impedance spectroscopy techniques were applied for the electrochemical characterization of the modified electrode in the presence of [Fe(CN)6]<sub>3-/4-</sub> as a redox probe. The system showed an exponential response in a concentration range from 50 pM to 500 nM of pirimicarb with a sensitivity of 5.74 × 10 <sub>-3</sub> µA pM<sub>-1</sub> mm<sub>-2</sub> , and the Limit of detection (LOD) is 10.30 pM.
Event details
- Event
- TC19 Symposium: Metrology on Environmental Instrumentation and Measurements (EnvIMEKO 2022)
- Technical Committee
- TC19
- envimeko.lemans.2022@gmail.com
- Place
- La Mans, FRANCE
- Time
- 2 June 2022 - 3 June 2022
- Website
- https://easychair.org/cfp/EnvIMEKO_2022