Traceable ammonia quantification and metrological uncertainty evaluation in a shock tube

D. Zhu, S. Agarwal, B. Shu, R. Fernandes, Z. Qu

Abstract

This work emphasizes on the development of an ultra-rapid spectrally resolved tunable diode laser absorption spectroscopy (TDLAS)-based spectrometer with a scan frequency of 40 kHz for dynamic NH3 quantification in a shock tube. Thanks to the high laser scan frequency, the NH3 mole fraction at various stages during the dynamic process can be quantified. Besides, considering lacking metrology in shock tubes for dynamic studies, we comprehensively evaluated the uncertainty sources and budgets of thermodynamic parameters and species concentration based on Guide to the expression of uncertainty in measurements (GUM). The established metrological uncertainty evaluation method for shock tube experiments can be beneficial to provide traceable and high-quality data, which is vital for dynamic studies as well as chemical kinetic modelling.

Keywords
Ammonia, TDLAS, Shock tube, Uncertainty budget
Download
IMEKO-TC24-2023-01.pdf
DOI
10.21014/tc24-2023.01
IMEKO TC
TC24 - Chemical Measurements

Event details

Event
IMEKO TC8, TC11 and TC24 Conference
Technical Committee
TC24
Email
info_ijconf_madeira2023@relacre.pt
Place
Madeira, PORTUGAL
Time
11 October 2023 - 13 October 2023
Website
http://imekojconfmadeira2023.lnec.pt/index.html

Back to the proceedings