Skip to main content

Status and performance of the LNE-Cnam Fabry-Perot refractometer

A. Rezki, Z. Silvestri, D. Bentouati, J.-P. Wallerand, C. Guianvarc’h, M. Himbert, P. Otal

Abstract

We present the status of the single-cavity Fabry-Perot interferometer developed at the LNE-Cnam laboratory used for thermodynamic pressure measurements within the range of 100 Pa to 100 kPa. After characterizing the intrinsic parameters of the refractometer, this optical sensor is used to measure the refractivity of nitrogen. Then, using the Lorentz-Lorenz equation and knowing the refractive virial coefficients at 532 nm, it is possible to deduce its density. Measuring the temperature of the gas makes it possible to determine its pressure using an equation of state. </BR>Once the temperature and pressure stability of the gas inside the optical sensor have been achieved at sub-mK and mPa levels, respectively, the expanded uncertainty of the sensor is evaluated to be [(?? ? ?? )^? + (?? × ??^−? ∙ ?)^? ]^(?/?). In order to further decrease this uncertainty, several approaches and solutions are given, leading to more accurate and reliable pressure measurements. The developed optical cavity operates as a high-resolution pressure sensor with an objective of complementing and eventually replacing conventional pressure standards, such as the force-balanced piston gauge and capacitance diaphragm gauge, that are based on the classical definition of pressure.

Download
IMEKO-TC16-2023-08.pdf
DOI
10.21014/tc16-2023.08
IMEKO TC
TC16 - Pressure and Vacuum Measurement

Event details

Event
TC16 Conference 2023
Technical Committee
TC16
Email
ccm2023@list.nist.gov
Place
Washington DC, UNITED STATES
Time
15 May 2023 - 19 May 2023
Website
https://www.nist.gov/news-events/events/2023/05/2023-ccm-imeko-tc16-7th-international-conference-pressure-and-vacuum

Back to the proceedings