Skip to main content

ERRORS DUE TO MAGNETIC EFFECTS IN 1 kg PRIMARY MASS COMPARATORS

Richard Davis, Josette Coarasa

Abstract

A revised draft of OIML R111 has been approved recently. Among the many notable changes to the previous edition are specific requirements for both the magnetic susceptibility and residual magnetization of weights used in legal metrology. The highest class of such weights is E1, for which a 1 kg weight has a manufacturing tolerance of ± 0.5 mg. It therefore follows that better magnetic properties might be needed for weights having mass uncertainties substantially better than E1 tolerances. The work presented below shows how this problem was approached at the BIPM and supports our conclusion that alloys meeting the new Class E1 specifications for 1 kg weights have magnetic properties sufficient for the needs of the BIPM.

Download
IMEKO-TC3-2005-029u.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
Force, Mass and Torque Measurements
Linked User
Dirk Röske
Technical Committee
TC3
Email
imeko@nlab.org
Place
Cairo, EGYPT
Time
19 February 2005 - 23 February 2005
Website
http://www.nlab.org/imeko/

Back to the proceedings