Skip to main content

FORCE TRACEABILITY MEASUREMENTS TO ACHIEVE 0.005 % BEST MEASUREMENT CAPABILITY

Hans W. Werner, Emil Harreus

Abstract

The increasing accuracy of force transducers and their calibration had two major effects: the chance to better extract the capability of a reference standard machine (compared to a formerly mixed evaluation of machine plus transducer) and on the other hand the need to build up reference standard machines capable to calibrate those high precision transducers. Using the experience of designing force transducers, of constructing calibration machines and of calibrating force within the German Calibration Service (DKD) since 1977 HBM decided to take the challenge. Two new dead-weight force reference standard machines of 2.5 kN and 25 kN have been built up and their best measurement capability of significantly better than 0.01 % has been verified by transfer transducers from HBM´s product range.

Download
IMEKO-TC3-2002-019.pdf
DOI
-
IMEKO TC
TC3 - Measurement of Force, Mass, Torque, and Gravity

Event details

Event
Joint International Conference on Force, Mass, Torque, Hardness and Civil Engineering Metrology in the age of globalization
Technical Committee
TC3
Place
Celle, GERMANY
Time
24 September 2002 - 26 September 2002

Back to the proceedings