Phase Error Determination for the Wideband Resistive Voltage Divider
Abstract
A new design of resistive voltage divider with series-parallel connection has been built. A selfcalibration method has also been proposed to determine the phase angle errors of the resistive voltage divider. The phase angle displacement has been measured by means of sampling with a high-speed digitizer. The measurement has been done at frequencies from 10 kHz to 100 kHz to calibrate two RVDs with ratio of 100:1 and the measurement results are presented.