Skip to main content

Comparison of pVTt Methods Gas Flow Prover

Meng Tao, Wang Chi

Abstract

We compared 20 m3 and 2 m3 facilities of pVTt methods gas flow primary prover (pVTt prover) in Flow lab of National Institute of Metrology of China (NIM). We used sonic critical venturi nozzle (sonic nozzle) as transfer standard. The flow points of comparison are about 100 nm3/h and 150 nm3/h. We completed four comparison tests whose results are considered acceptable, whereas we found there was a little systematic error between the two facilities.

Keywords
Comparison, pVTt methods, Nozzle
Download
IMEKO-TC9-2004-021.pdf
DOI
-
IMEKO TC
TC9 - Flow Measurement

Event details

Event
tc9-2004

Back to the proceedings