THE EFFECT OF ADJUSTMENT ON MULTIFUNCTION INSTRUMENTS FOR ELECTRICAL METROLOGY: CASE EXAMPLES

Giuseppe La Paglia, Gian Carlo Bosco, Marco Lanzillotti,  Luca Callegaro
Abstract:
Top-class multifunction instruments (multimeters and calibrators) implement in their firmware adjustment procedures requiring a minimal set of adjustment standards. The small set of standards' values provided have effects on the (usually very broad) ranges of values and frequencies which the instrument can measure. How errors and uncertainties in the adjustment of specific points affect measurements in such broad ranges has been experimentally investigated in top-class multimeters and calibrators; most interesting results have been observed in the ‘ac voltage’ function; a number of examples are reported.
Download:
IMEKO-TC4-2008-055.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
Exploring New Frontiers of Instrumentation and Methods for Electrical and Electronic Measurements
Title:
XVIth IMEKO TC4 International Symposium on Electrical Measurements and Instrumentation (together with 13th IMEKO TC4 Workshop on ADC Modelling and Testing)
Place:
Florence, ITALY
Time:
22 September 2008 - 24 September 2008