ESTABLISHMENT OF AN AUTOMATED 10 V JOSEPHSON SERIES ARRAY VOLTAGE STANDARDAT NATIONAL PHYSICAL LABORATORY, INDIA

V. N. Ojha, Shiv Kumar Jaiswal and Sudhir Kumar Sharma
Abstract:
In the present paper, we discuss the integration, characterization and operation of the recently established an automated 10 volt Josephson Series Array Voltage Standard (JSAVS) at National Physical Laboratory, India (NPLI). Furthermore, we discuss the precision calibration of the Zener reference standard alongwith the uncertainty budget. The combined standard uncertainty of the Zener reference standard calibration at 10V for a Single Point Measurement (SPM) [1] is ± 25 nV at k=1.
Download:
IMEKO-TC4-2008-068.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
Exploring New Frontiers of Instrumentation and Methods for Electrical and Electronic Measurements
Title:
XVIth IMEKO TC4 International Symposium on Electrical Measurements and Instrumentation (together with 13th IMEKO TC4 Workshop on ADC Modelling and Testing)
Place:
Florence, ITALY
Time:
22 September 2008 - 24 September 2008