GLUCOSE UPTAKE RATE UNCERTAINTY IN A FED BATCH BIOREACTOR

G.Sassi, M.Bernocco
Abstract:
Glucose uptake rate is a typical quantity for bioreactor non invasive monitoring, which is candidate as metabolic assay for tissue engineering. Volumetric glucose uptake rate enables to compare bioreactor performances. Fed Batch reactors are typical in tissue engineering applications. This work aims to the identification of the metrological bottlenecks of an uptake rate measurement in a bioreactor for human cells; it offers a first metrological characterization of the quantities for a typical bioreactor protocol. Glucose uptake rate and volumetric one are measurable at 6% and 3% uncertainty respectively. The enhancement of bioreactor management protocols and design, and volume measurement may lead under 1% uncertainty.
Keywords:
uncertainty, uptake rate, bioreactor, glucose, hMSC
Download:
IMEKO-WC-2012-TC13-O1.pdf
DOI:
-
Event details
Event name:
XX IMEKO World Congress
Title:

Metrology for Green Growth

Place:
Busan, REPUBLIC of KOREA
Time:
09 September 2012 - 12 September 2012