OPTIMIZATION OF CALIBRAION INTERVAL AND ADJUSTMENT LIMIT OF MESUREMENT DEVICE WITH STOCHASTICALLY DIFFUSIVE PROPERTY

K. Shirono, H. Tanaka, M. Shiro, K. Ehara
Abstract:
A calibration and an adjustment of measurement device are keys to realize a metrological traceability. (Calibration means just checking without adjustment in this paper.) Although the traditional online quality engineering can be one of applications to determine calibration interval and adjustment limit in terms of cost, approximations used in the traditional quality engineering are too rough to take stochastic behaviour of a bias in a device into consideration. In this paper, we propose an optimum calibration interval and an optimum adjustment limit for a measurement device with a bias in accordance with Brownian dynamics concerning a long-term cost.
Keywords:
online quality engineering, Brownian dynamics, stochastic differential equation, calibration interval, adjustment limit
Download:
IMEKO-WC-2012-TC21-O19.pdf
DOI:
-
Event details
Event name:
XX IMEKO World Congress
Title:

Metrology for Green Growth

Place:
Busan, REPUBLIC of KOREA
Time:
09 September 2012 - 12 September 2012