EFFECT OF TEMPORARY TEMPERATURE CHANGE ON STABILITY OF WIRE-WOUND RESISTANCE STANDARDS |
|---|
| J. C. Ryu, Y. T. Park, K. S. Han, K. M. Yu |
- Abstract:
- Irreversible resistance changes of Thomas type 1 ohm resistors according to temporary temperature change is about 0.03 ppm/y in average by reported results of PTB, IEN and NIST, but the behavior did not occur in NML 1 ohm resistors under the same condition. By qualitative analysis based on thermal stress and strain, it is considered that a major element of the behavior is thermal stress by different thermal expansions of coils and mandrels.
- Keywords:
- irreversible resistance changes, Thomas type resistors, NML type resistors, thermal stress
- Download:
- IMEKO-WC-2000-TC12-P317.pdf
- DOI:
- -
- Event details
- Event name:
- XVI IMEKO World Congress
- Title:
Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future
- Place:
- Vienna, AUSTRIA
- Time: