DEVELOPMENT OF 0,015 TO 2 BAR GAUGE PRESSURE STANDARD

D. Zvizdic, L. Grgec Bermanec
Abstract:
This paper presents calibration system in the gauge pressure range of 0,015 bar to 2 bar developed in Laboratory for Process Measurement (LPM) at Faculty of Mechanical Engineering and Naval Architecture (FSB) in Zagreb. Design of the system, theoretical basis, main parameters influencing the effective pressure and the model for estimation of measurement uncertainty are described. The scope of the system is presented through pressure sensor calibration with results presented in tabular form.
Keywords:
pressure, calibration, uncertainty estimation
Download:
IMEKO-WC-2000-TC16-P444.pdf
DOI:
-
Event details
Event name:
XVI IMEKO World Congress
Title:

Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future

Place:
Vienna, AUSTRIA
Time:
25 September 2000 - 28 September 2000