COMPACT, WIDE RANGE ROTATING VISCOMETER

H. Stabinger
Abstract:
A rotary viscometer using an eddy current brake for the measurement of momentum. The instrument covers a range of 0.5 to 30000 mPa·s without changing the measuring cell. The resolution is better than 0.1% and the accuracy 0.2%. A Peltier thermostat controls the measuring temperature from -20°C to 100°C with 0.01°C accuracy. Fast response and a short measuring time are the benefits of this instrument.
Keywords:
rotary viscometer, momentum measurement, eddy current brake
Download:
IMEKO-WC-2000-ST-P474.pdf
DOI:
-
Event details
Event name:
XVI IMEKO World Congress
Title:

Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future

Place:
Vienna, AUSTRIA