Abstract:
Transfer measurements from the standard of direct voltage to the standard of alternative voltage is widely used in measurement and instrumentation. Well known method is based on the usage of thermo converter of voltage. There are some drawbacks of this method such as instability of error, hand mode of calibration and inertia of reproduction process. The new method of reproduction AC (AV) value proposed by usage of inertia’s properties of links of the parallel-resonance circuit to transfer measurements from the standard of direct current to the standard of alternative current. The paper illustrates the mechanism of action of AC-calibrator based on proposed method of reproduction, where AC signal is coming through L-inductor without its changes.
Download:
IMEKO-TC4-2010-060.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
Instrumentation for the Information and Communication Technology Era
Title:
XVIIth IMEKO TC4 Symposium & 15th International Workshop on ADC Modelling and Testing (together with 3rd IMEKO TC19 Symposium)
Place:
Kosice, SLOVAKIA
Time:
08 September 2010 - 10 September 2010