INCREASED ACCURACY OF MEASUREMENT OF HIGH-SENSITIVITY MM-RANGE RADIOMETRIC EQUIPMENT INTENDED FOR MEDICAL-BIOLOGICAL APPLICATION |
|---|
| Yury Skripnik, Aleksandr Krasuk, Vyacheslav Manoylov, Sergey Peregudov, Aleksey Yanenko |
- Abstract:
- The paper considers the methods of improving the accuracy of measurements of low-intensity electro-magnetic radiation of the bio-objects in mm-range carried out with the use of high-sensitivity radiometric equipment with the reference noise generator included into its structure. The main sources of uncertainty of measurement are deter-mined. It is shown that with the help of a reference noise generator having the power comparable with radiation power of bio-objects, it is possible to decrease essentially the uncertainty of measurement and to register accurately the peculiarities of radiation spectrum of the living objects.
- Keywords:
- radiometric system, uncertainty of measurement, reference noise generator