A PIV EXAMINATION ON THE PERFORMANCE OF ULTRASONIC HEAT METER RELATED TO INSTALLATION

Wang Sun, Luhai Wang, Jiming Yang, Yujian Zhu
Abstract:
An experimental investigation was carried out on the performance of an ultrasonic heat meter with interference of non-uniformity from the upstream installation of a ball valve, which it is partly opened. It was found that the existence of the ball valve may under-report the flow, especially when the flow rate is low. From the measured PIV results, it can be found that the velocity distribution is obviously non-uniform inside the meter, and the change is remarkable when the inlet is added with a ball valve. The time resolved PIV results also demonstrate that the flow field becomes unsteady when slow down the flow rate, which might be responsible for the measurement deviation.
Keywords:
PIV, ultrasonic heat meter, ball valve
Download:
IMEKO-TC9-2013-053.pdf
DOI:
-
Event details
IMEKO TC:
TC9
Event name:
FLOMEKO 2013
Title:

16th International Flow Measurement Conference

Place:
Paris, FRANCE
Time:
24 September 2013 - 26 September 2013