ACCURACY ENHANCEMENT OF A COMBINED V/Z CLAMP-ON ULTRASONIC FLOW |
|---|
| Sejong Chun, Byung-Ro Yoon, Woong Kang, Hyu-Sang Kwon |
- Abstract:
- Ultrasonic flow metering with clamp-on type transducers is a promising method when portability in flow metering is necessary. Nevertheless, the clamp-on ultrasonic flow meter has low measurement accuracy as flow speed in the pipe decreases. If the two types of clamp-on ultrasonic flow paths, namely, Z-shaped and V-shaped ultrasonic paths, are combined, measurement accuracy can be increased because the metering output becomes robust. In the present study, the measurement accuracy of the combined clamp-on ultrasonic flow meter was within ±2 % at 50 m3/h, which was smaller than those of the Z-shaped and the V-shaped clamp-on ultrasonic flow meters at the same flow rate. This can provide an advantage for water resource management because a general guideline for the water resource management is less than ±2 %.
- Keywords:
- clamp-on ultrasonic flow meter, V-shaped, Z-shaped
- Download:
- IMEKO-TC9-2013-083.pdf
- DOI:
- -
- Event details
- IMEKO TC:
- TC9
- Event name:
- FLOMEKO 2013
- Title:
16th International Flow Measurement Conference
- Place:
- Paris, FRANCE
- Time: