THE PROBLEM OF ACCURACY IN ROUGHNESS MEASUREMENT WITH THE USE OF THE FORM MEASURING MACHINE TALYROND 365 |
|---|
| Marta Wisniewska, Sabina Zebrowska-Lucyk |
- Abstract:
- Accuracy of roughness parameters measured with Talyrond 365 was examined with comparison to the results received with the Form Talysurf PGI 830. The research has shown that there are several limitations of measurement method and software as well, while measuring with the Talyrond 365. Also, there are significant differences between results acquired with machines under comparison. The reasons for observed dissimilarities were discussed.
- Keywords:
- form tester, accuracy, roughness
- Download:
- IMEKO-TC14-2013-71.pdf
- DOI:
- -
- Event details
- IMEKO TC:
- TC14
- Event name:
- TC14 ISMQC 2013
- Title:
11th International Symposium on Measurement and Quality Control
- Place:
- Cracow and Kielce, POLAND
- Time:
- 11 September 2013 - 13 September 2013