A PRACTICAL METHOD TO DETECT THE REPEATABILITY OF A GRAVIMETRIC FLOW TEST RIG BY USING THE TIMING ERROR

Moritz Leopoldo Cordova Murillo, Thomas Lederer
Abstract:
The present paper presents an improved method for the diverter error determination method according to the international recommendation ISO4185. By extending this test it is possible to make a qualitative test of the overall status of a gravimetric ?ow test rig. The proposed method has been successfully applied to results of diverter tests of the large water ?ow calibration facilities of the PTB in Braunschweig and Berlin. It is shown that by applying this method also the repeatability of the ?ow meter used, and of the test bench alone can be estimated.
Keywords:
diverter error, ISO 4185, repeatability, gravimetric flow test rig
Download:
IMEKO-TC9-2013-013.pdf
DOI:
-
Event details
IMEKO TC:
TC9
Event name:
FLOMEKO 2013
Title:

16th International Flow Measurement Conference

Place:
Paris, FRANCE
Time:
24 September 2013 - 26 September 2013