Experimental investigation on dynamical performances of a novel fiber-optic pressure sensor for pulmonary ventilation

Luigi Battista, Andrea Scorza, Giulia Lupi, Salvatore Andrea Sciuto
Abstract:
A new optical fiber differential pressure sensor has been proposed for neonatal pulmonary ventilation and an experimental investigation on its dynamical metrological performances is here reported. The proposed measurement system has been preliminarily tested by monitoring the mouth pressure during simple respiratory tests in healthy patients and during mechanical ventilation performed by means of a infant ventilator typically used in the neonatal intensive care units; results show that the output of the transducer is consistent with that measured by means of a reference sensor.
Keywords:
Dynamical measurement, Fiber-optic sensor, Pressure sensor, mechanical ventilation
Download:
IMEKO-TC4-2014-216.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
TC4 Symposium 2014
Title:

20th IMEKO TC4 Symposium on Measurements of Electrical Quantities (together with 18th TC4 International Workshop on ADC and DCA Modeling and Testing, IWADC)
"Research on Electrical and Electronic Measurement for the Economic Upturn"

Place:
Benevento, ITALY
Time:
15 September 2014 - 17 September 2014