Design, stability analysis and uncertainty contribution of a voltage divider designed for a phase meter

Stanislav Maslan, Martin Sira, Vera Nováková Zachovalová
Abstract:
Today uncertainties of a phase difference of two sinusoidal signals is commonly about 10 μrad for non coherent measurements and signal amplitudes up to several volts. Increasing voltage range of a phase meter requires special considerations. The construction of a special voltage divider intended for precise phase difference measurement is presented in this paper. The uncertainty contribution of the voltage divider was determined by repeated measurement and an Allan variance calculation. The result showed out the voltage range of the phase meter increased substantially with very small addition to the total uncertainty of the phase difference.
Keywords:
phase measurement, measurement uncertainty, voltage dividers, stability analysis
Download:
IMEKO-TC4-2014-192.pdf
DOI:
-
Event details
IMEKO TC:
TC4
Event name:
TC4 Symposium 2014
Title:

20th IMEKO TC4 Symposium on Measurements of Electrical Quantities (together with 18th TC4 International Workshop on ADC and DCA Modeling and Testing, IWADC)
"Research on Electrical and Electronic Measurement for the Economic Upturn"

Place:
Benevento, ITALY
Time:
15 September 2014 - 17 September 2014