A NEW DESIGN OF A 5,4 MN BUILD-UP SYSTEM |
---|
Gerhard Schulder, Ulrich Kolwinski, Daniel Schwind |
- Abstract:
- The main purpose of the build-up system is the use in verification measurements of force standard machines up to 5MN or even higher capacity. The special design of the described build-up system allows the use of nine force transfer standards either individually, in groups of three transducers or all transducers together in three groups of three. A 9-channel digital measuring amplifier (GTM VNDigitizer 3.0) enables the user to get the readout of all transducers simultaneously.
- Keywords:
- build-up, force-transducer, calibration
- Download:
- PWC-2006-TC3-009u.pdf
- DOI:
- -
- Event details
- Event name:
- XVIII IMEKO World Congress
- Title:
Metrology for a Sustainable Development
- Place:
- Rio de Janeiro, BRAZIL
- Time:
- 17 September 2006 - 22 September 2006