BEHAVIOUR OF HIGH VALUE STANDARD RESISTORS VERSUS RELATIVE HUMIDITY: FIRST EXPERIMENTAL RESULTS |
|---|
| F. Galliana, P.P. Capra, E. Gasparotto |
- Abstract:
- In this paper we report the first experimental results in the characterization of high value standard resistors versus relative humidity. For this aim we adopted a digital multimeter (DMM) based measurement method, already used for characterization of high value resistors versus temperature and applied voltage and to participate in the CCEM-K2 comparison on high resistance. For the realization of the environment with specific values of relative humidity we realized a thermostatic enclosure in which it is possible to set the desired humidity value in the range (10 ÷ 90)% with a long term instability of the humidity up to ± 1%. The preliminary experimental results of characterization suggest us that the value assigned to a high value standard resistor is affected not by the humidity during the calibration period, but by the humidity with which it was maintained several days or weeks before calibration.
- Keywords:
- high value standard resistor, relative humidity, thermostatic enclosure
- Download:
- IMEKO-TC4-2001-005.pdf
- DOI:
- -
- Event details
- IMEKO TC:
- TC4
- Event name:
- TC4 Symposium 2001
- Title:
11th IMEKO TC4 Symposium on Trends in Electrical Measurements and Instrumentation (together with 6th IMEKO TC4 Workshop on ADC Modelling and Testing)
- Place:
- Lisbon, PORTUGAL
- Time:
- 13 September 2001 - 14 September 2001