USING THE NOSE-TO-NOSE SAMPLER CALIBRATION METHOD IN PULSE METROLOGY |
|---|
| D. R. Larson, N. G. Paulter |
- Abstract:
- We present our analysis of the nose-to-nose (ntn) method for use as an accurate sampler calibration method. The variations in the measurement of the sampler impulse response using the ntn method are described and the assertion that the kick-out pulse is identical to the sampler impulse response is assessed. Temperature effects on the ntn method are also examined. Finally, the effect of uncertainties in the ntn calibration method on the uncertainties in reported pulse parameters is examined.
- Keywords:
- sampler calibration, nose-to-nose, kick-out pulse, offset voltage
- Download:
- IMEKO-TC4-2001-137.pdf
- DOI:
- -
- Event details
- IMEKO TC:
- TC4
- Event name:
- TC4 Symposium 2001
- Title:
11th IMEKO TC4 Symposium on Trends in Electrical Measurements and Instrumentation (together with 6th IMEKO TC4 Workshop on ADC Modelling and Testing)
- Place:
- Lisbon, PORTUGAL
- Time:
- 13 September 2001 - 14 September 2001