CALIBRATION OF FIBER OPTIC POWER METER AT 1625 nm |
|---|
| Seung Kwan Kim, Kee Suk Hong, Ki-Lyong Jeong, Yong Seok Kim |
- Abstract:
- Fiber optic power responsivity scale at 1625 nm is newly established to support optical fiber communication industry. Using the absolute scale at reference level of -10 dBm and the linearity measurements, a fiber optic power meter is calibrated at 1625 nm from -60 dBm to 0 dBm. The uncertainty is less than 0.7 % (k = 2) at -10 dBm and about 1 % in the full range.
- Keywords:
- fiber optic power, responsivity, calibration, fiber optic power meter
- Download:
- IMEKO-WC-2015-TC2-018.pdf
- DOI:
- -
- Event details
- Event name:
- XXI IMEKO World Congress
- Title:
Measurement in Research and Industry
- Place:
- Prague, CZECH REPUBLIC
- Time:
- 30 August 2015 - 04 September 2015