DESIGN, DEVELOPMENT AND ESTABLISHMENT OF 1 kN FORCE CALIBRATION MACHINE IN FORCE LABORATORY OF TSE

Bulent Aydemir, Sinan Fank, Cemal Vatan, Dervis Anil, Alican Kuyrukluyildiz
Abstract:
A new dead weight type force calibration machine with a 1 kN loading capacity both in tension and compression is installed both in force laboratory of UME and TSE. This paper describes the preliminary results of the performance of the newly developed 1 kN dead weight force machine at the Force Laboratory. The machine is designed and manufactured by UME. The performance test results show that relative reproducibility error with rotation (b) ± 5,5 × 10-3 and relative repeatability error without rotation (b’) ± 3,5 × 10-3 for a range of 0,2 kN to 1 kN of a new 1 kN machine of TSE.
Keywords:
force calibration machine, dead weight, performance test
Download:
IMEKO-WC-2015-TC3-029.pdf
DOI:
-
Event details
Event name:
XXI IMEKO World Congress
Title:

Measurement in Research and Industry

Place:
Prague, CZECH REPUBLIC
Time:
30 August 2015 - 04 September 2015