THE TEMPERATURE CHANGE DURING MEASUREMENT VIA MASS COMPARATOR WITH AUTOMATIC LOAD ALTERNATOR

Feng-yu Yang, Sheau-shi Pan, Kuang-Yih Tsuei
Abstract:
We changed a new 10-kg mass comparator with automatic load alternator at CMS (Center for Measurement Standards) last year to replace the older one which was out of function. During installation, the temperature issue comes out to affect the measurement results under the same environment of laboratory as usual. In this paper, we will discuss the abnormal phenomenon caused by the thermal effect.
Keywords:
mass comparator, automatic load alternator, thermal effect