MEASUREMENT OF THE CURRENT TRANSFER RATIO FOR CURRENT SHUNTS |
|---|
| Pavel F. Baranov, Edvard I. Tsimbalist, Valeriy N. Borikov, Diana G. Soltanova |
- Abstract:
- Current shunts are widely used in metrological practice. The shunt is required to determine the value of its resistance on DC and AC currents to determine the transfer ratio error in the wide frequency band in process calibration. The paper describes a procedure for measuring the complex impedance of transfer ratio of current shunts in the frequency band using the lock-in amplifier with the differential input. The experimental results of the proposed procedure are described.
- Keywords:
- current shunt, lock-in amplifier, transfer ratio, calibration, complex impedance
- Download:
- IMEKO-WC-2015-TC4-102.pdf
- DOI:
- -
- Event details
- Event name:
- XXI IMEKO World Congress
- Title:
Measurement in Research and Industry
- Place:
- Prague, CZECH REPUBLIC
- Time:
- 30 August 2015 - 04 September 2015