Abstract:
To answer the need for traceability in micro flow area for medical industry, Flow Laboratory of Metrology LIPI in cooperation with Physics Department of University of Indonesia, developed a compact standard based on piston displacement concept. The standard consists of glass syringe and linear actuator controlled by Field Programmable Gate Array (FPGA). The volume generation uncertainty is 0.6% to 0.7% at 6.5 mL to 13.4 mL. Flow-rate uncertainty is 1.6% to 11.6% at 90.3 mL/hr to 1177.1 mL/hr.