- Abstract:
- This study analyzes an end-shorting effect on electromagnetic flowmeters with short distance between connecting flanges around about 5D or more. The Bevir’s formulation is extended to a conducting pipe wall. Through the evaluation of the Green function satisfying the corresponding boundary conditions, the author’s previous formula is reviced.