DEVELOPMENT THE PRIMARY VELOCITY STANDARD FOR THE CALIBRATION OF LDA SYSTEM

Wu Jian, Nikolay Semin
Abstract:
Air velocity measurements are required to have the measurement traceability to the SI unit of length and time. This can be realised through the setup of a primary velocity standard. NMC has designed and developed the primary velocity standard using the spinning disc for the calibration of an LDA system. The system is capable of measuring the velocity from 0.2 m/s to 40 m/s with the measurement uncertainty of 0.1% to 0.4% of reading.
Keywords:
Laser Doppler Anemometry (LDA), air velocity, primary standards, uncertainty evaluation
Download:
IMEKO-WC-2015-TC9-229.pdf
DOI:
-
Event details
Event name:
XXI IMEKO World Congress
Title:

Measurement in Research and Industry

Place:
Prague, CZECH REPUBLIC
Time:
30 August 2015 - 04 September 2015