THE COMPARISON OF A MICROWAVE BASED BIOENERGY MOISTURE MEASUREMENT INSTRUMENT AGAINST THE LOSS-ON-DRYING METHOD |
|---|
| Petri Österberg, Martti Heinonen, Maija Ojanen-Saloranta, Anssi Mäkynen |
- Abstract:
- This article compares the performance of a microwave based moisture measurement instrument against the conventional moisture measurement reference method Loss-on-Drying (LoD). Six different biomaterials were measured at three moisture content levels with both measurement methods. After instrument calibrations, the difference and variation of the measurement results and the repeatability of the instruments for parallel samples were estimated. Good agreement between the measurement methods was achieved with careful microwave instrument calibration.
- Keywords:
- moisture measurement, bioenergy, microwave, oven drying, LoD
- Download:
- IMEKO-WC-2015-TC24-458.pdf
- DOI:
- -
- Event details
- Event name:
- XXI IMEKO World Congress
- Title:
Measurement in Research and Industry
- Place:
- Prague, CZECH REPUBLIC
- Time:
- 30 August 2015 - 04 September 2015