Investigation of the Calibration Capabilities of a 1 kN·m And 10 kN·m Reference Calibration Machine

F. Crößmann
Abstract:
To expand the capacity of its calibration laboratory, Hottinger Baldwin Messtechnik GmbH is planning to use two new reference calibration machines in the range up to 1 kN·m and 10 kN·m respectively. In addition to the theoretical consideration of the individual elements contributing to measurement uncertainty, measurement data is also compared using the En value as a demonstration of the target measurement uncertainty of 0,02% (k = 2).
Keywords:
torque, calibration machine, measurement uncertainty
Download:
IMEKO-TC3-2017-014.pdf
DOI:
-