UNCERTAINTY EVALUATION OF THE NANOINDENTATION SYSTEM |
|---|
| Jiong-Shiun Hsu, Chi-Shen Chang, Yu-Yi Su, Jeah-Sheng Wu, Sheau-Shi Pan |
- Abstract:
- In this paper, the uncertainty of nanoindentation system was evaluated. The laser interferometer and electronic balance were respectively adopted to calibrate the quantities of displacement and load of nanoindeantion system. The results reveal that the relative expanded uncertainty of indeantion hardness, UH, is 8.1%. The relative expanded uncertainity of reduced modulus, UEr, is 5.4%.
- Keywords:
- uncertainty, nanoindentation system
- Download:
- PWC-2006-TC5-009u.pdf
- DOI:
- -
- Event details
- Event name:
- XVIII IMEKO World Congress
- Title:
Metrology for a Sustainable Development
- Place:
- Rio de Janeiro, BRAZIL
- Time: