PRECISE DETERMINATION OF GE DETECTOR EFFICIENCY CURVE FOR OBTAINING ACTIVITIES IN RADIONUCLIDES GAMMA-EMITTERS |
---|
Jose U. Delgado, Maria Cd. M. de Almeida, Roberto Poledna |
- Abstract:
- An efficiency curve obtained in a gamma-ray spectrometric system using a high purity germanium detector and radioactive standards of 133Ba, 152Eu and 166mHo in an energy range of 50 to 1400 keV was established for metrological purposes. To do this, the efficiency response of the detector was used considering 29 different energies. This allowed calibrating various gammaemitter radioactive point sources, within of a 50 kBq to 550 MBq activity range, with expanded uncertainties below 2.5% at two-sigma confidence level (k=2). The obtained activity value for 22Na was validated by sum-peak coincidence method.
- Keywords:
- efficiency curves, gamma-emitters, Ge detector
- Download:
- PWC-2006-TC7-011u.pdf
- DOI:
- -
- Event details
- Event name:
- XVIII IMEKO World Congress
- Title:
Metrology for a Sustainable Development
- Place:
- Rio de Janeiro, BRAZIL
- Time:
- 17 September 2006 - 22 September 2006