THE TRIPLE POINT OF XENON: A CANDIDATE FIXED-POINT REFERENCE TEMPERATURE |
|---|
| K.D. Hill, A.G. Steele |
- Abstract:
- The triple point of xenon has recently been shown to be a suitable fixed point for incorporation into the next revision of the International Temperature Scale as a means to reduce the non-uniqueness in the important 84 K to 273 K range. We summarize the results, which illustrate that the xenon triple point itself is highly reproducible, with a standard deviation of 48 µK for the eight melts of this study, and a total realization uncertainty of just 76 µK.
- Keywords:
- international temperature scale, fixed point
- Download:
- PWC-2006-TC12-001u.pdf
- DOI:
- -
- Event details
- Event name:
- XVIII IMEKO World Congress
- Title:
Metrology for a Sustainable Development
- Place:
- Rio de Janeiro, BRAZIL
- Time: