THE TRIPLE POINT OF XENON: A CANDIDATE FIXED-POINT REFERENCE TEMPERATURE

K.D. Hill, A.G. Steele
Abstract:
The triple point of xenon has recently been shown to be a suitable fixed point for incorporation into the next revision of the International Temperature Scale as a means to reduce the non-uniqueness in the important 84 K to 273 K range. We summarize the results, which illustrate that the xenon triple point itself is highly reproducible, with a standard deviation of 48 µK for the eight melts of this study, and a total realization uncertainty of just 76 µK.
Keywords:
international temperature scale, fixed point
Download:
PWC-2006-TC12-001u.pdf
DOI:
-
Event details
Event name:
XVIII IMEKO World Congress
Title:

Metrology for a Sustainable Development

Place:
Rio de Janeiro, BRAZIL
Time:
17 September 2006 - 22 September 2006