INTERNAL COMPARISON BETWEEN A REFERENCE TYPE TORQUE CALIBRATION MACHINE AND A 10 N·m DEADWEIGHT TORQUE STANDARD MACHINE

A. Nishino, M. Hamaji, K. Ogushi
Abstract:
A reference torque calibration machine (RTCM) has been developed to validate the calibration of efficient torque transducers. In this study, the calibration and measurement capability (CMC) of the RTCM was evaluated based on the technical guideline. We also compared the RTCM and a 10 N·m deadweight torque standard machine at the National Metrology Institute of Japan (NMIJ) using a torque measuring device with a nominal capacity of 1 N·m.
Keywords:
torque; calibration; reference type torque calibration machine; internal comparison; uncertainty