LOW FREQUENCY PRIMARY VIBRATION CALIBRATION USING A MULTI-COMPONENT SHAKER

N. Yan, L. Klaus, T. Bruns
Abstract:
The multi-component vibration calibration device at PTB consists of a triaxial shaker which is suitable and convenient for calibrating multi-axial seismometers. Three laser vibrometers are furthermore utilised as references. The frequency range was further extended for a European research project. Instead of a closed-loop control system limiting the vibration frequency down to 0.4 Hz, a signal generator is now used to provide the excitation voltage to the amplifier of the shaker in open-loop mode. This excites the shaker yielding vibrations with significantly lower frequencies. The first test measurements with a seismometer were carried out. The results demonstrate that a calibration down to 10 mHz is easily achievable.
Keywords:
primary vibration calibration; seismometer; multi-component shaker; low frequency
Download:
IMEKO-TC22-2022-019.pdf
DOI:
10.21014/tc22-2022.019
Event details
IMEKO TC:
TC22
Event name:
TC22 Conference 2022
Title:

5th TC22 Conference on Vibration Measurement (together with the 24th TC3 Conference on the Measurement of Force, Mass and Torque, the 14th TC5 Conference on the Measurement of Hardness, and the 6th TC16 Conference on Pressure and Vacuum Measurement)

Place:
Cavtat-Dubrovnik, CROATIA
Time:
11 October 2022 - 13 October 2022