Skip to main content

Proceedings

9762 papers found

Eulalia Balestrieri, Pasquale Daponte, Luca De Vito, Francesco Picariello, Sergio Rapuano, Ioan Tudosa

Experimental Assessment of a Novel CS-based Acquisition Method for ECG Signals in IoMT

The paper presents the results obtained from a preliminary experimental assessment of a novel method which relies on Compressed Sampling (CS) of electrocardiographic (ECG) signal. The mathematical model for CS-based acquisition of ECG signals and its feasibility for practical implementation for an Internet-of-Medical Things (IoMT) system are described. Herein, the presented experimental results demonstrate the robustness and convenience of the proposed method.

tc4-2019 · TC4 · 2019 · Download
E. Balestrieri, P. Daponte, L. De Vito, F. Picariello, S. Rapuano, I. Tudosa

DAC testing: recent research directions

Digital to Analog Converters (DACs) represent essential components in many applications. During the years, their performances have been improved hand in hand with the increasing need to accurately test them. Due to the large number of different DAC applications, several different specifications have been used to quantify the effective performance of these converters in a particular context. To unambiguously clarify DAC terms, definitions and test methods the IEEE Std. 1658 has been developed and is currently under revision. The paper presents an overview of the last research work dealing...

tc4-2019 · TC4 · 2019 · Download
Xiaoyuan Wu, Yan Yang, Zongtao Chi

Design of Small Capacitance Standards Based on Analytical and Simulation Approaches

In the paper, analytical model and simulation are adopted to design of small capacitance standards based on fused silica. The small capacitance standards is designed with Kelvin equipotential protection structure. The stray capacitance introduced by the air gap between the capacitance eletrode and the Kelvin equipotential protection electrode(guard electrode) will affect the small capacitance value. The analytical model of the capacitor is established and the Schwarz-Christoffel transformation is used to analyze the stray capacitance. By adjusting the size of the air gap between the central...

tc4-2019 · TC4 · 2019 · Download
Jing Gu, Yan Yang, Zhaosheng Teng

A High Isolation Analogue Switch and its Application in Digital Sampling Bridge

This paper describes a high-isolation analogue switching device to replace the mechanical reed switches used in precision digital sampling bridges. The switching speed of the designed analogue switch is much higher than that of the mechanical switches, and the Off-Isolation is good enough at frequencies between 1 kHz and 100 kHz. A method of applying the designed analogue switch to an automatic source balanced single digital sampling bridge is given, the theoretical measurement accuracy of the given bridge can reach ppm level with faster measurement speed. Finally, the actual Off-Isolation...

tc4-2019 · TC4 · 2019 · Download
Platon Sovilj, Vladimir Vujičić, Zoran Mitrović, Dragan Pejić, Bojan Vujičić, Đorđe Novaković

Metrology, Electrical Energy Markets and Engineering Education at University of Novi Sad (Serbia)

This paper begins with presentation of the main facts about University of Novi Sad (Serbia) and its Faculty of Technical Sciences. Important results of metrology in electrical engineering from past to nowadays, at Faculty of Technical Sciences (University of Novi Sad), are consequently described. The role of the Centre of Metrology is recognized as the basis for cooperation between university and industry in the field of metrology. All these parameters enabled ELEMEND – EU ERASMUS+ project for electrical energy markets and engineering education which is being implemented at University of...

tc4-2019 · TC4 · 2019 · Download
Norbert Rogge, Shan Lin, Christian Rothleitner, Suren Vasilyan

Excitation frequency dependent deviations during the „velocity mode” of Bl measurements in the Planck-Balance

The Planck-Balance is a table-top device that utilizes the concept of a Kibble balance. In contrast to a standard electromagnetic force compensated balance, the force factor B l is not determined by calibration with mass pieces. Instead of this static calibration, the Planck-Balance applies a sinusoidal motion on the coil and determines the B l based on measurements of induced voltage, displacement and frequency. In order to obtain valid force calibration for the weighing experiment, one has to ensure that the determined factor does not depend on the excitation frequency of the coil. A...

tc4-2019 · TC4 · 2019 · Download
Zengmin Wang, Honghui Li, Yuan Gao, Yunfeng Lu, Qing He

Applications of Programmable Josephson Voltage Standard on Magnetic Measurements

This paper describes the progress of magnetic applications on Programmable Josephson Voltage Standard (PJVS) at NIM. In past few years, we developed a PJVS based magnetic flux measurement system and continued improving the magnetic flux and the mutual inductance measurement accuracy. This paper introduces the automatic flux and mutual inductance measurement system that we developed. Experimental results show that the type A uncertainty of the measurement of mutual inductances of 577 mH is less than 2 ppm, which demonstrates the feasibility and reliability of the flux and mutual inductance...

tc4-2019 · TC4 · 2019 · Download
Y. A. Sanmamed, J. R. Salinas, J. Díaz de Aguilar, F. García-Lagos, R. Caballero

Temperature influence on the frequency response of the Keysight 3458A digital multimeter

This paper describes the procedure and experimental set-up used at CEM to determine the temperature influence on the frequency response of two Keysight 3458A digital multimeters (DMMs) in the Direct Current Volt (DCV) sampling mode. The DMMs were used to digitalize an AC input signal of 0.8 V and 1 kHz provided by an AC source.

tc4-2019 · TC4 · 2019 · Download
Wenjing Liu, Haiyang Zhang, Bo Gao, Changzhao Pan, Dongxu Han, Jiangfeng Hu, Yaonao Song, Hui Chen, Xiangjie Kong, Ercang Luo, Laurent Pitre

Comparsion of different kinds of frequency scanning methods used at low temperature

A novel single pressure refractive index gas thermometry (SPRIGT) has been collaboratively developing by the TIPC-CAS in China and the LNE-Cnam in France. In the SPRIGT, microwave resonance frequencies are used to determine the refractive index of a working gas in a quasi-spherical resonator at a single pressure and temperature. Then an unknown temperature can be determined by comparing the refractive index with that at a reference temperature. Usually, the resonance frequencies were obtained by fitting the S21 scattering parameters with respect to the scanning frequencies, which are...

tc4-2019 · TC4 · 2019 · Download
Leonidas Niyonkuru, Gerd Vandersteen, Leo Van Biesen

Power spectrum measurement of high frequency angle modulated digital signal from output power spectrum of a digital divider

This paper shows how to measure high frequency angle modulated digital signals using a relatively low frequency spectrum analyzer and a digital frequency divider. The mathematical proof is based on comparison of the signal spectrum before and after the frequency divider. The spectral scaling of angle modulated digital signal by the divider is predictable mathematically implying that the spectrum at the input of the divider can be deduced from the spectrum at the digital output of the frequency divider. The proposed method was confirmed by measurement.

tc4-2019 · TC4 · 2019 · Download