READING OF THERMOMETERS USING A VIDEO CAMERA

V. Batagelj, J. Bojkovski, I. Pušnik
Abstract:
In the proposed paper a possibility of automation of calibration of liquidin- glass thermometers is discussed. Temperature readings are taken with the use of a measuring system that consists of a standard calibration facility, the video camera, and the personal computer that executes the program for image processing and analyses. Once the reading of temperature is extracted from an acquired image, this value can be passed to any other data processing, which includes the comparison with the values from the reference thermometer and determination of total uncertainty. Data achieved in this way can be directly used to prepare a calibration certificate.
Keywords:
calibration automation, image processing, liquid-in-glass thermometers
Download:
IMEKO-WC-2000-UNC-P495.pdf
DOI:
-
Event details
Event name:
XVI IMEKO World Congress
Title:

Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future

Place:
Vienna, AUSTRIA
Time: