- Abstract:
- The paper presents design and basic metrological properties of three thermal AC voltage standards with nominal input voltages 1.5 V, 3 V and 6 V. Each standard is composed of a range resistor connected in series with a single junction thermal converter (SJTC). The AC-DC transfer differences of these standards are calculated in frequency range from 10 kHz to 1 MHz.
- Keywords:
- AC voltage standard, AC-DC transfer, thermal voltage converter