PTB AND INRIM HIGH FORCE INTERCOMPARISON UP TO 9 MN

C. Ferrero, C. Marinari, R. Kumme, W. Herte
Abstract:
The present paper describes the long term stability of the main metrological characteristics of the two reference transducers of 3 MN and 9 MN together with the relevant uncertainties evaluated from the measurements carried out by INRIM and PTB by bilateral comparisons. In June 2004, the new INRIM – 3 MN build-up system (BU), the 9 MN and 5 MN transducers were also intercompared, by using the PTB 2 MN DWM and the 16.5 MN Hydraulic Multiplication Machines (HMM). The main results are reported. Furthermore long term investigations carried out over 20 years and the work obtained within a EUROMET project are described.
Keywords:
force standard machine, long term stability, intercomparison
Download:
PWC-2006-TC3-002u.pdf
DOI:
-
Event details
Event name:
XVIII IMEKO World Congress
Title:

Metrology for a Sustainable Development

Place:
Rio de Janeiro, BRAZIL
Time:
17 September 2006 - 22 September 2006